Fluke Calibration 5730A Operator's Manual page 181

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

SUBNETMASK?
Description
Sequential command. Returns Ethernet subnet mask for LAN communication.
Parameter
None
Response
String
SUBNETMASK?
Example
Returns 255.255.254.0 if the subnet mask was previously set to this value.
*TST?
Sequential command, ignored if not in remote.
Initiates a series of self-tests, then returns a "0" for pass or a "1" for fail. If any faults are
Description
detected, they are logged into the fault queue where they can be read by the FAULT?
query.
Parameter
None
Response
(Integer) 0 (for Pass) or 1 (for Fail)
UNCERT?
Sequential command.
Description
Returns the calculated maximum uncertainty of the Calibrator output according to the
selected calibration interval.
Parameter
None
1.
(Float) Uncertainty of the Calibrator output (-1.0 if no specification is available)
Response
2.
(String Units for response 1 (PPM, PCT, V, A, OHM, etc.)
3.
(Integer) The specification interval in days.
VOUT?
Sequential command.
Description
Returns the output amplitude expressed in volts if the currently selected units are dBm.
Parameter
None
1.
(Float) Output amplitude
Response
2.
(String) V, A, OHM, units for response 1
3.
(Float) Frequency
Assuming a value of 10 dBm, 10 kHz is currently active on the main output terminals,
Example
VOUT? returns: "2.4494897E+00,V,1.0000E+04"
Table 6-5. Commands (cont.)
Remote Commands and Syntax
6
Commands
6-47