Fluke Calibration 5730A Operator's Manual page 154

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Operators Manual
CAL_CLST?
Description
Parameter
Response
Example
CAL_CONF
Description
Parameter
CAL_CONF?
Description
Parameter
Response
CAL_CONST?
Description
Parameter
Response
Example
CAL_COUNT?
Description
Response
Parameter
Example
6-20
Table 6-5. Commands (cont.)
Sequential command.
Returns a list of names values of a particular group of calibration constants.
CAL
(Returns active calibration constants)
CHECK
(Returns calibration check constants)
PREV
(Returns previous calibration constants)
(String)
"<EOL>
<total number of constants> <EOL>
<name> <value> <EOL>
<name> <value> <EOL>
(cont)
"
CAL_CLST?, CAL
Could return:
"
424
D3P, 3.9817876E+02
(Continues for about 400 pairs of names and values.)
"
Sequential command.
Sets the calibration specifications to a confidence level of either 99% or 95%.
CONF95 or CONF99: Sets 95% or 99% specifications.
Sequential command.
Returns the current calibration confidence level.
None
CONF95 (indicates 95% specifications)a or CONF99 (indicates 99% specifications)
Sequential command.
Returns the value of a particular calibration constant from a particular group of constants.
1. CAL
(From active calibration constants)
CHECK
(From calibration check constants)
PREV
(From previous calibration constants)
2. Symbolic name of the calibration constant desired (see Appendix D).
(Float) The value of the constant
CAL_CONST? CHECK, KV6
Could return: 6.5000010E+00
Sequential command. Returns the number of calibration constants that have been saved
at the end of a calibration procedure (except for DC Zeros)
integer
None
CAL_COUNT?
Returns 34 if calibration constants have been saved 34 times.