2.

Installation

Select the Dionex SRD-10 Installation Site
The Dionex SRD-10 must be installed within 50.8 cm (20 in) of the suppressor,
due to the length of the Dionex SRD-10 inlet tubing. It can be installed so as to
allow the regenerant or external water supply to exit the suppressor and flow
through the Dionex SRD-10 to waste.
Use the hook-and-loop strips on the rear of the Dionex SRD-10 to mount it on the
laboratory bench, the side of the IC system, or other flat surface. Any orientation
is acceptable.
Install the Dionex SRD-10
1. Turn off the power to the IC system (including the pump, detector,
autosampler, and chromatography compartment).
2. Connect the Dionex SRD-10 inlet tubing (labeled
(see
Figure
View A
Figure 1. Thermo Scientific Dionex SRD-10 Suppressor Regenerant Detector
Doc. 065355-04 05/18
Dionex SRD-10 Installation Instructions
1, View A) to the
REGEN OUT
Inlet Tubing
TTL Output
USB Port
TO SRS/AES, REGEN OUT
port on the suppressor.
View B
Waste
Line
Hook-and-loop
strip (1 of 3)
)
3