Dionex SRD-10 Installation Instructions
3. Place the free end of the Dionex SRD-10
in a waste container below the level of the workbench on which the IC system
is installed.
4. Remove the backing from the three hook-and-loop strips on the Dionex SRD-
10 (see
Figure
5. While holding the Dionex SRD-10 in the preferred orientation, press it firmly
onto the mounting surface.
6. Refer to the table below to identify the TTL input that provides on/off control
of the pump motor.
Product
Dionex GP50, GS50, ICS-3000 SP, ICS-5000 SP*,
IP25, or IS25
Dionex ICS-1000, ICS-1500, or ICS-2000
Dionex ICS-3000 DP or ICS-5000 DP*
Dionex ICS-1100, ICS-1600, or ICS-2100
* Analytical pump only
7. Locate the appropriate connector plug in the Dionex SRD-10 Ship Kit. Follow
these basic steps to connect the TTL:
a. Connect the active wire (red) and ground wire (black) from the Dionex
SRD-10 to the connector plug.
To attach a wire to the plug, strip the end of the wire (if it is not already
stripped), insert it into the plug, and use a screwdriver to tighten the
locking screw. If necessary, multiple ground wires can be attached to a
single TTL input/output ground pin.
When attaching wires to the connector plug, be careful not to allow stray
strands of wire to short to the adjoining position on the connector.
b. Plug the Dionex SRD-10 connector plug into the connector on the rear
panel of the integrated IC system or the pump module.
4
1, View B).
line (see
Figure
WASTE
TTL Input
TTL Input 1
TTL Input 3
Pump 1: TTL Input 2
Pump 2: TTL Input 3
Any available TTL input
Doc. 065355-04 05/18
1, View A)