4.7.2 Programming / modifying the Special / Multi wavelengths meth-
ods
Note
For multi wavelength methods, you can use the method numbers 2001 to
2050. All special methods can also be selected in the method list of the con-
centration mode.
The creation of a user-defined method is done in the following steps:
Enter the general method data
Method number, method name, unit etc.
Enter the wavelengths for absorbance measurements (A
Minimum 1, maximum 10
Define the procedure variables (K
Procedure variables are used to take into account any influence quanti-
ties that cannot be measured by the photometer.
The values for these procedure variables have to be entered for all mea-
surements with the method, e.g. the temperature or pH value.
Enter the formula to calculate the measurement result
Enter the formula with which you want to calculate the measurement
result in the formula editor.
Enter an additional condition (optional)
Conditions are used to check the measurement result for validity.
The condition is entered with the formula editor.
) (optional)
X
Operation
)
x nm