Operation With The 12 V Adapter Ada 12V - wtw photoLab 6100 VIS Operating Manual

Hide thumbs Also See for photoLab 6100 VIS:
Table of Contents

Advertisement

®
photoLab
6100 VIS
Safety instructions
Operating time with
battery
Example
Technical data
ADA 12V
ba75847e06
06/2013

3.4.6 Operation with the 12 V adapter ADA 12V

With the ADA 12V you can operate the photoLab
eter on the move and independent of the local power supply.
To do so, a 12 V power supply such as a commercial 12 V portable power
source or a 12 V car battery is required.
12 V power supply unit
(e.g. portable power
source or car battery )
For operation with an external battery, follow the safety instructions of the
battery.
Make sure the power source is suitable for operation of the spectrophotome-
ter (see technical data of the power source and technical data of the spectro-
photometer).
The maximum operating time depends on various factors:
 Battery (e.g. nominal capacity, condition, age)
 Operating mode of the spectrophotometer (e.g. frequency of measure-
ments)
 Photometer (instrument type)
Operating time with a type 12 V / 19 Ah battery in optimum condition: approx.
6.5 h
Note
The spectrophotometer consumes electricity even while it is in standby
mode.
We recommend to disconnect the ADA 12V if you do not use the photometer
during battery operation.
Cable length
max. voltage
max. current
ADA 12V
photoLab
2 m
12 V
8 A
Commissioning
®
6100 VIS spectrophotom-
®
6100 VIS
25