AQA in measured
value
documentation
ba75847d06
06/2013
4.15
Analytical quality assurance (AQA)
4.15.1
General information
The target of the analytical quality assurance (AQA) is to secure correct and
precise measurement results.
Note
Settings for AQA checks are only available for users of the user group,
administrator.
Every registered user can carry out the AQA check (see also section 4.16.1).
Analytical quality assurance (AQA) can be carried out in two steps indepen-
dent of each other:
AQA1: Monitoring of the photometer
AQA2: Monitoring of the total system.
It comprises the photometer, the used test, the accessories and the user's
way of working.
The monitoring includes a check procedure that has to be successfully
repeated by the user within a certain period (AQA interval).
Note
The AQA monitoring is not active in the delivery condition.
All values that are measured after a passed check and within the AQA inter-
val are given the addition Protocol ID in the measured value documentation.
This addition is used to identify the relevant AQA inspection record.
Operation