photoLab
6100 VIS
ba75847d06
06/2013
4.18
Photometer information ([Info])
The following photometer information is displayed:
Photometer designation
Version number of the meter software/method data
Hardware version
Series number of the meter
Registered user
Hardware status (for service purposes)
Memory status
<HOME>
[Info]
Info
Model designation:
Serial number:
Software/methods version:
Build:
Hardware version:
Hardware status:
Lamp counter
System test
Filter test
Wavelength calibration
Free internal memory space
Registered user
4.19
Lamp counter
The photometer counts the operating hours of the lamp. The information on
the operating hours of the lamp is given in the Info menu.
The number quoted there corresponds to the number of measurements.
After the lamp was exchanged (see section 5.1) you should reset the lamp
counter (see section 4.17).
The meter information and result
16.04.07 9:52
®
of the self-test are displayed and
photoLab
6100 VIS
07440001
can be printed.
1.30-WTW-1.60
04/03/09 11:57
0-
FF 00000000
12
✓
✓
✓
✓
✓
Operation
165