Operation
56

4.11 Administrating user-defined methods

4.11.1 Options for method administration

User-defined programs can be
 entered
 read out
 erased.
To store user-defined programs (methods) in the meter, determine a calibra-
tion line for your method yourself and transmit the data of this calibration line
®
to the pHotoFlex
STD.
Administration of the method data can be carried out in the LSdata software
(see LSdata operating manual) or with the aid of a terminal program ( see
section 4.11.2).
You need the following data in any case:
Data
Program No:
Model:
Cell [mm]:
Wave length [nm]:
MRB (measuring range start):
MRE (measuring range end):
E0 (reagent blank value):
Slope:
Resolution:
Citation form:
Unit:
Blank required:
(No/Yes)
Standard adjustment possible:
(No/Yes)
Standard adjustment required:
(No/Yes)
Possible entries/examples
900 ... 949
Any name (max. 5 characters)
16 or 28
436, 517, 557, 594, 610 or 690
e.g. 0.05
e.g. 8.00
e.g. 0.0
e.g. 1.0
0.0001, 0.001, 0.01, 0.1, 1, 2, 5,
10 or 100
e.g. PO4-P (max. 9 characters)
e.g. mg/l (max. 7 characters)
0 or 1
0 or 1
0 or 1
ba75958e07
®
pHotoFlex
STD
11/2020