D Series Syringe Pumps
Section 7 Modbus Configuration
Table 7-5 Coils
Register
Address
00001
0
00002
1
00003
2
00004
3
00005
4
00006
5
00007
6
00008
7
00009
8
00010
9
00011
10
00012
11
00013
12
00014
13
00015
14
00016
15
00017
16
00018
17
00019
18
00020
19
00021
20
00022
21
00023
22
00024
23
00025
24
00026
25
00027
26
00028
27
00029
28
00030
29
00031
30
00032
31
00033
32
00034
33
00035
34
00036
35
00037
36
7-4
Type
Size
Description
R/W
1 BIT
0: Stop Pump A 1: Run Pump A
R/W
1 BIT
0: Stop Pump B 1: Run Pump B
R/W
1 BIT
0: Stop Pump C 1: Run Pump C
R/W
1 BIT
0: Stop Pump D 1: Run Pump D
R/W
1 BIT
Link Run/Stop operation of multi-pump pair AB & multi-pump pair CD 0: Disable 1:Enable
R/W
1 BIT
0: Stop Pump A if in Refill 1: Run Pump A in Refill
R/W
1 BIT
0: Stop Pump B if in Refill 1: Run Pump B in Refill
R/W
1 BIT
0: Stop Pump C if in Refill 1: Run Pump C in Refill
R/W
1 BIT
0: Stop Pump D if in Refill 1: Run Pump D in Refill
R/W
1 BIT
1: Pump A in Constant Pressure mode
R/W
1 BIT
1: Pump B in Constant Pressure mode
R/W
1 BIT
1: Pump C in Constant Pressure mode
R/W
1 BIT
1: Pump D in Constant Pressure mode
R/W
1 BIT
1: Pump A in Constant Flow mode
R/W
1 BIT
1: Pump B in Constant Flow mode
R/W
1 BIT
1: Pump C in Constant Flow mode
R/W
1 BIT
1: Pump D in Constant Flow mode
R/W
1 BIT
1: Pump A & Pump B in Independent mode
R/W
1 BIT
1: Pump A & Pump B in Continuous Flow Constant Flow mode
R/W
1 BIT
1: Pump A & Pump B in Continuous Flow Constant Pressure mode
R/W
1 BIT
1: Pump A & Pump B in Modifier addition mode
R/W
1 BIT
1: Pump A & Pump B & Pump C in Continuous Modifier addition mode
R/W
1 BIT
1: Pump C & Pump D in Independent mode
R/W
1 BIT
1: Pump C & Pump D in Continuous Flow Constant Flow mode
R/W
1 BIT
1: Pump C & Pump D in Continuous Flow Constant Pressure mode
R/W
1 BIT
0:Multi-pump pair AB in Delivery mode 1: Multi-pump pair AB in Receive mode
R/W
1 BIT
0: Multi-pump pair AB in Low Press operation 1: Multi-pump pair AB in Normal Press operation
0: Use Fast (coarser) pressure match control for multi-pump pair AB
R/W
1 BIT
1: Use Normal (finer) pressure match control for multi-pump pair AB
R/W
1 BIT
0: Multi-pump pair CD in Delivery mode 1: Multi-pump pair CD in Receive mode
0: Multi-pump pair CD in Low Press operation 1: Put multi-pump pair CD in Normal Press opera-
R/W
1 BIT
tion
0: Use Fast (coarser) pressure match control for multi-pump pair CD
R/W
1 BIT
1: Use Normal (finer) pressure match control for multi-pump pair CD
R/W
1 BIT
1: Activates the automatic rapid pressurization cycle (Constant Flow mode only) for Pump A
R/W
1 BIT
1: Activates the automatic rapid pressurization cycle (Constant Flow mode only) for Pump B
R/W
1 BIT
1: Activates the automatic rapid pressurization cycle (Constant Flow mode only) for Pump C
R/W
1 BIT
1: Activates the automatic rapid pressurization cycle (Constant Flow mode only) for Pump D
R
1 BIT
1: Pump A is Equilibrating
R
1 BIT
1: Pump B is Equilibrating