Electric valves, which provide the best positive valve closure, are
most commonly used in liquefied gas applications. Electric valves
require the pump controller to be equipped with the valve driver
board.
E30D
Note
Electrical valves can only be used with pumps A and B.
Dual system maximum flow rates under optimal conditions are
approximately 65% of the max flow rate for each pump model.
Table B-2 Continuous Flow Technical Specifica-
tions-Electric Valves:
Pressure fluctuation at
switch over
Minimum
Maximum system back
pressure
Maximum flow rate (ml/min)
Liquids: 65% of the single
pump maximum rate
(excluding E1000 systems).
Liquefied gases
Temperature range
Wetted materials in valve package:
Tubing and fittings
62-1247-113
3.44 bar, at system back pressures from 6.9
bar to the single pump maximum. Higher
fluctuation occurs at pressures below 6.9
bar.
3.5 bar
The single pump maximum. Valves rated to
2068 bar
E30D
14.3
45% of the single pump maximum rate. Cyl-
inder cooling jackets should be used to
obtain this rate.
5 to 40° C
SS316, gold