D Series Syringe Pumps
Section 1 Introduction
Figure 1-12 Menu 1
1-22
3. Check the upper left corner of the controller screen. The
current pump mode will appear in a two-letter abbrevia-
tion, e.g. CF for constant flow. This will be followed by a
lowercase letter indicating the current pump, e.g. "a" indi-
cates that pump A is the current pump. The current pump
is the one for which parameters are being set.
a. If a pump other than pump A is currently selected:
In the lower right corner of the screen, directly over
softkey D, are the words "SELECT PUMP."
Press
> 2 > A

MENU

automatically switch to the run screen, showing "a" in
the upper left corner.
4. Press
to display Menu 1.
MENU
1. UNITS
2. SELECT PUMP
3. REFILL
MORE
5. Press 1 for
.
UNITS
6. In the Units menu, select the desired units to display for pres-
sure. For example, press 3
units.
7. In the Units menu, select the desired units to display for flow.
For example, press
The first line of the display will show the selected units.
8. Press
to return to Menu 1.
(D)
PREVIOUS
9. Press
, to exit Menu 1.
(D)
RETURN
10. Press
to set the pump mode to constant flow.
CONST FLOW
CFa will be displayed in the upper left corner of the screen.
11. Press
(A)
FLOW RATE
flash on the display. Use the numeric keys to enter a flow
rate of 10 ml/min and press
Note
If you make an error, press
12. Press
. Observe the flow rate displayed on the first line.
RUN
After a few moments, the setpoint and flow rate display
should match.
13. Once the setpoint and flow rate match, press
a. If three or fewer pumps are connected, you will be
prompted to press A to stop pump A, B to stop pump B,
C to stop pump C, or D to stop all pumps.
to select pump A. The display will
4. POWER FAILURE [STOP]
5. SYSTEM RESET
6. DISPLAY CONTRAST
RETURN
to select PSI for the pressure
5 to select ML/MIN for the flow units.
. The words
ENTER FLOW RATE
.
ENTER
to delete it.
CLEAR ENTRY
should
.
STOP

Hide quick links: