D Series Syringe Pumps
Section 3 Basic Programming and Operation
3.14 Analog Flow Rate and
Volume Output
Options

3.14.1 Voltage Analog Output

This output can also be:
4-20 mA OUTPUT
ANALOG OUTOUT
ELECTRIC VALVES
12-CH OUTPUT
25-pin cable
Figure 3-14 Analog voltage output connection, rear panel
3-22
If two pumps are used with the controller, the second control wire
should be connected to terminal 2, under DIGITAL INPUT.
If three pumps are used with the controller, the third control wire
should be connected to terminal 3, under DIGITAL INPUT.
If four pumps are used with the controller, the fourth control wire
should be connected to terminal 4, under DIGITAL INPUT.
Note
For the 4-20ma option, the negative current wire is connected
to DIGITAL GROUND and the positive should be connected to
DIGITAL INPUT
The analog output option provides for analog monitoring of the
syringe pump flow rate. If two or less pump modules are con-
nected to the controller, pump volume delivered can also be mon-
itored via analog output. These outputs are often used with
analog based plant or process monitoring equipment.
Output voltage for pressure is located on the back panel of the
pump module via standard banana jacks.
The output connection is the female 25-pin Sub-D connector
located on the rear panel of the controller, under 4-20mA Output.
DIGITAL I/O
4-20mA OUTPUT
USB / ETHERNET
Controller rear panel
RS-232 / RS-485
Voltage analog outputs are pre-installed at the factory. This
option has a total of four outputs for monitoring flow rate with
selectable ranges from 0 to 5V, -5 to +5V, and 0 to 10V. The
default range at installation is 0 to 10V. Each output can be set to
a different range.
The top of the selected range is the same maximum value
entered for
MAX FLOW
Section 3.8.7 of this manual. For example: if the output range is
when setting
, as described in
LIMITS

Hide quick links: