Serial Control Check List; Controller Setup; Restore Defaults (If Desired); Change Defaults - Teledyne D series Installation And Operation Manual

Table of Contents

Advertisement

8.5 Serial Control Check
List

8.6 Controller Setup

8.6.1 Restore Defaults (if
desired)

8.6.2 Change Defaults

In order to build and operate a custom program for Isco syringe
pumps, you must have the following:
1. Sample program (Isco LabView toolkit)*
2. LabView Compiler (purchased separately)
3. Connection cable(s)
4. 1 controller: cable 480-7996-00
5. 2 controllers: cable 68-1020-198
6. 3 to 7 controllers: cables 68-1020-198 and 68-1020-180.
7. Source code (serial commands in computer language for
DASNET conversion)
8. DASNET serial driver
*The toolkit comes with directories for two different versions of
LabView. Copy the .zip files onto your computer.
The files within each directory are as follows:
\labview8.5
CONTROL.LLB
DASNET.DLL
DATA.LLB
ISCO1.LLB
ISCO2.LLB
ISCO3.LLB
MULTPUMP.LLB
PANEL.VI
README.DOC
SERIALIO.LLB
Once the system is properly connected, turn the pump, con-
troller(s), and computer on. Then use the following procedure to
place your controller(s) in serial control mode.
Press
>
(A) and select
MENU
MORE
Factory controller default settings are baud rate 9600 and unit
ID #6.
To restore defaults, press
Resetting a system erases all programs and user settings.
1. Press MENU > MORE (A) > 1. Serial Option to enter the
serial option setup screen.
2. Select the communication baud rate (between 300 and
115.2K) by pressing
desired baud rate.
Note
All units in the network must be set to the SAME baud rate.
D Series Syringe Pumps
Section 8 Serial Interface
\labview5-8.2
CONTROL.LLB
DASNET.DLL
DATA.LLB
ISCO1.LLB
ISCO2.LLB
ISCO3.LLB
MULTPUMP.LLB
PANEL.VI
README.DOC
SERIALIO.LLB
(1).
SERIAL
, then
MENU
SYSTEM RESET
(A) until you reach the
NEXT BAUD
(5).
8-5