Section 7 Modbus Configuration

7.1.2 Modbus TCP/IP

7.1.3 Modbus Configuration
Options
7-2
The modbus TCP/IP protocol implementation uses an
ethernet connection. The ethernet connections are located on the
optional DB25 connector labeled USB/Ethernet. Use optional
cable (PN 60-1244-488) and coupler (PN 14-9905-200) to connect
to your Ethernet.
Pin on DB 25 Casing
2
14
15
17
22
23
To access the Modbus configuration settings, press:
(
)
MORE
(
)
MORE
(
)
MORE
Table 7-1 Modbus TCP/IP Configuration Options
Item
IP Address
Internet protocol address of device
(Default 192.168.1.200)
Subnet Mask
Subnet mask (Default 255.255.255.0)
Gateway
Gateway (Default 0.0.0.0)
Port
TCP port (Default 502)
Table 7-2 Modbus RTU Configuration Options
Item
Baud Rate
Slave ID
Ethernet
Transmit +
Transmit -
Receive +
Receive -
Vgnd
Vgnd
Modbus Options
1. Modbus RTU
2. Modbus TCP
Description
Value
300, 1200, 2400, 4800, 9600,
Communication Speed (Default
19200, 38400, 57600, 115200
1-247
Device address (Default 1)
RJ-45
1
2
3
6
4+5
7+8
Description
19200)