Seal Selection Chart - Teledyne D series Installation And Operation Manual

Table of Contents

Advertisement

D Series Syringe Pumps
Section 2 Fluid System Connections and Accessories
Seal Type
Max
Pressure
GENERAL/SFE
0-689.5
CARBON
bar
DIOXIDE
This is our standard seal. It is good for most applications, particularly those using organic solvents,
such as: LC, SFC, and SFE.
HIGH
0-689.5
TEMPERATURE
bar
This seal is best suited for applications which require temperatures between 50°C and 100°C. Its
chemical compatibility is similar to that of the general seal.
LOW
0-137.9
PRESSURE
bar
This seal may provide better sealing at lower pressure. Its chemical compatibility is similar to that of
the general seal.
AQUEOUS
0-689.5
bar
This seal has better wetting properties, making it a good choice for aqueous solutions.
It is also the best choice for electrochemical detection.
Note: This seal requires a special break-in procedure before installation. See Section 9.9.2.
AMMONIA (NH3)
0-275.8
NITRIC ACID
bar
This is the only seal recommended for ammonia.
2-20
Table 2-7 Seal Selection Chart
Seal
65D/DM
Description
Black, PTFE
202-9096-08
graphite filled
Black, PTFE graphite fiber
reinforced high temperature
compound
Black, PTFE graphite filled
single point contact
White-translucent, ultra-high
molecular weight polyethyl-
ene
White, virgin PTFE
Note
Call the factory for seal availability for the 30D pump.
100DX/DM
260D
202-9090-75 202-9091-06 202-9091-56 upper seal
202-9090-76 202-9091-09 202-9093-56 upper seal
202-9092-75 202-9092-06 202-9092-56
202-9090-77 202-9094-06 202-9094-56
202-9090-78 202-9091-07 202-9091-57 upper seal
500D/
1000D
500HPX
202-9990-25
lower seal
202-9990-23
202-9990-28
lower seal
202-9990-27
N/A
N/A
202-9990-26
lower seal
202-9990-24

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

30d100dx65d260d500d1000d ... Show all