D Series Syringe Pumps
Section 3 Basic Programming and Operation

3.5.1 Serial Option

Figure 3-4 Serial option menu

3.5.2 Pump Status

3.5.3 External Control

3.5.4 Multiple Pumps

3.5.5 Total Volume Reset

Figure 3-5 Total volume reset
3-6
The serial option menu allows you to set the baud rate and the
unit identification number.
BAUD RATE = 1200
UNIT # ID = 6
NEXT ID#
NEXT BAUD
A
Use softkey A, NEXT BAUD, to scroll through the available baud
rates. These are: 300, 1200, 2400, 4800, 9600, 19.2K, 38.4K,
57.6K, and 115.2K.
Use softkey B, NEXT ID#, to scroll through the available ID
numbers. These are 1-7. Six is the default, as this is the unit
identifier for the Teledyne Isco LabView™ software.
For a complete discussion of serial control, refer to Section 8 in
this manual.
This option displays the controller software revision, and model
of connected pump(s).
The first line displays the revision of the software.
Lines 2, 3, and 4 display the type of pump connected to the A, B,
C, and D pump connectors, respectively.
This screen is also momentarily displayed automatically each
time the pump controller is switched on.
The pump pressure or flow rate operation can be controlled
externally with an analog voltage. Complete instructions for this
feature are provided in Sections 3.12 through 3.12.5.
One controller can control up to four pumps at once, either
together or independent of each other.
Complete information about using the
vided in Section 3.9 and Section 5.6.
This option resets the total volume display of multi-pump pair
AB or multi-pump pair CD to zero when operating in continuous
flow or modifier mode.
1. AB PUMPS
2. CD PUMPS
3. AB AND CD PUMPS
PREVIOUS
B
C
MULTI PUMP
D
feature is pro-