KROHNE OPTIWAVE 5200 C/F Handbook page 65

Non-contactradar (fmcw) level meter ordistance, leveland volume measurementof liquids,pastesand slurries
Hide thumbs Also See for OPTIWAVE 5200 C/F:
Table of Contents

Advertisement

OPTIWAVE 5200 C/F
Cable gland
Cable entry capacity (terminal)
Input and output
Output signal
Resolution
Temperature drift
Error signal
Approvals and certification
CE
Vibration resistance
Explosion protection
ATEX (pending) KEMA
10ATEX0036 X
IECEx (pending) IECEx KEM
10.0006 X
cFMus - Dual Seal-approved
(pending)
NEPSI (pending)
03/2012 - 4001904901 - HB OPTIWAVE 5200 R01 en
Standard: none
Options: M20×1.5 (cable diameter: 6...10 mm / 0.23...0.39¨); others are available on request
0.5...1.5 mm²
4...20 mA HART
or 3.8...20.5 mA acc. to NAMUR NE 43
®
±3 µA
Typically 50 ppm/K
High: 22 mA; Low: 3.6 mA acc. to NAMUR NE 43
This device fulfils the statutory requirements of the EC directives. The manufacturer certifies successful testing of
the product by applying the CE mark.
EN 60068-2-64
Metallic Horn: (DN200): 5 Hz to 100 Hz: 4g
PTFE or PP Wave Horn: 3.5 mm up to 8 Hz and 10 m/s²: 1g, 8.5 to 2000 Hz
II 1/2 G, 2 G Ex ia IIC T6...T2 Ga/Gb or Ex ia IIC T6...T2) Gb;
II 1/2 D, 2 D Ex ia IIIC T90°C Da/Db or Ex ia IIIC T90°C Db IP6X;
II 1/2 G, 2 G Ex d ia IIC T6...T3 (or T2) Ga/Gb or Ex d ia IIC T6...T3 (or T2) Gb;
II 1/2 D, 2 D Ex ia tb IIIC T90°C Da/Db or Ex ia tb IIC T90°C Db IP6X
Ex ia IIC T6...T3 (or T2) Ga/Gb or Ex ia IIC T6...T3 (or T2) Gb;
Ex ia IIIC T90°C Da/Db or Ex ia IIIC T90°C Db IP6X;
Ex d ia IIC T6...T3 (or T2) or Ex d ia IIC T6...T3 (or T2) Gb;
Ex ia tb IIIC T90°C Da/Db or Ex ia tb IIIC T90°C IP6X
NEC 500
XP-IS / Cl. I / Div. 1 / Gr. ABCD / T6;
DIP / Cl. II/III / Div. 1 / Gr. EFG / T6;
IS / Cl. I/II/III / Div. 1 / Gr. ABCDEFG / T6;
NI / Cl. I / Div. 2 / Gr. ABCD / T6
NEC 505
Cl. I / Zone 0 / AEx d [ia] / IIC / T6;
Cl. I / Zone 0 / AEx ia / IIC / T6;
Cl. I / Zone 2 / AEx nA [ia] / IIC / T6;
Hazardous (Classified) Locations, indoor/outdoor Type 4X and 6P, IP66, Dual Seal
CEC Section 18 (Zone ratings)
Cl. I, Zone 1, Ex d, IIC (Antenna: Zone 0), T6;
Cl. I, Zone 0, Ex ia, IIC, T6;
Cl. I, Zone 2, Ex nA, IIC, T6 DIP A21 IP66 TB 95°C
CEC Section 18 and Annex J (Division ratings)
Cl. I, Div. 1/2, Gr. ABCD; Cl. II, Gr. EFG; Cl. III, T6;
Ex ia IIC T2~T6 DIP A21 T
IP66;
A
Ex dia IIC T2~T6 DIP A21 T
IP66
A