ELECTRICAL CONNECTIONS
OPTIWAVE 5200 C/F
4
Figure 4-2: How to open the terminal compartment cover
• Turn the cover counterclockwise with a strap wrench.
• Remove the cover.
Figure 4-3: Procedure for electrical installation
Equipment needed:
• Small slotted tip screwdriver (not supplied)
CAUTION!
Make sure that the polarity of the wires is correct.
Procedure:
1 Do not disconnect the safety cord from the terminal compartment cover. Put the terminal compartment cover adjacent to the
housing.
2 Remove the connector from the circuit board.
3 Connect the electrical wires to the connector. Attach the connector to the circuit board. Tight- en the cable entry glands.
29
03/2012 - 4001904901 - HB OPTIWAVE 5200 R01 en
www.krohne.com