KROHNE OPTIWAVE 5200 C/F Handbook page 54

Non-contactradar (fmcw) level meter ordistance, leveland volume measurementof liquids,pastesand slurries
Hide thumbs Also See for OPTIWAVE 5200 C/F:
Table of Contents

Advertisement

OPERATION
6
Menu
No.
2.2.0
2.2.1
2.2.2
2.3.0
2.3.1
2.3.2
2.3.3
54
Function
Function description
Emp.Spec.Type
5
Use the average value for tanks
which contain fixed objects
only. Use the maximum value
for tanks which contain many
objects or moving objects.
Recording in progress
6
Reading in progress
Empty spectrum graph
7
Push [>] to make a selection
from the available spectra to
get the correct level signal, if
the tank is partially filled.
Do you want to save the
8
spectrum?
TESTS
The device serial number.
SET OUTPUT
This sets the current output to a test value
[mA]. The output will change to the given
value, independent of the measured value.
The current output will go back to the
measured value when the display goes back
to the menu level.
This starts the hardware test. Press [>]
DIAGNOSTIC
many times to show: the time of operation
(D1), temperature of the electronic
converter board (T1), loop current (I1), load
current (I2), voltage 5.6 V (V1), voltage on
capacitors (V2), voltage 3.3 V (V3), reset
counter (C1). If you press [>] again, the
display goes back to the menu level.
BASIC PARAM.
The basic parameters for
configuration of the device.
TANK HEIGHT
The distance from the flange face /
thread stop of the tank connection down
to the tank bottom.
BLOCK. DIST.
Blocking distance. The distance from the
flange to the top limit of the measuring
range (a zone given by the user where it is
not possible to measure). We recommend a
blocking
distance of 500 mm / 19.7¨ below the
antenna. If the distance is less than the
blocking distance, the device continues
to display the blocking distance.
TIME CONST.
Using this function, the device processes
several measurement readings to filter out
disturbances. Increasing the time constant
will
smoothen the integrated readings,
decreasing will roughen the
readings. s=seconds.
www.krohne.com
OPTIWAVE 5200 C/F
Selection list
Maximum, Average
Yes [>], No [ ]
3.5, 4, 6, 8, 10, 12, 14, 16,
18, 20 or 22 mA
min-max:
0...30 m / 0...98.4 ft
min: 100 mm / 13.9¨ or
2.3.6 ANTENNA.EXT
max: 2.3.1 TANK HEIGHT
min-max:
1.00...100.00 s
03/2012 - 4001904901 - HB OPTIWAVE 5200 R01 en
Default
?
3.5 mA
10.0 m /
32.8 ft
500 mm /
19.7¨
10.00 s