Usb Interface; Chassis Gnd; Ac Power Cord; Fuses - Newport 6100 User Manual

Laser diode and temperature controller
Table of Contents

Advertisement

System Operation
4.5.1

USB Interface

The instrument is designed to communicate with standard USB Host
interfaces. The connector on the rear panel is a standard USB-B (Full-Size,
Device).
4.5.2

Chassis GND

This 4 mm banana jack is connected to chassis ground. It is intended to be
used as an additional earth ground connection for the Model 6100's enclosure.
4.5.3

AC Power Cord

The Model 6100 can be operated on either 50 or 60 Hz mains power. The
instrument can be configured for operation at the following nominal AC line
voltages -- 100, 120, 220, 230, or 240 VAC. See section 3.6 for information
on properly configuring the Model 6100 for the available mains power.
The line cord supplied with each unit should be plugged only into a properly
grounded outlet to prevent electrical shock in the event of an internal short
circuit to the metal cabinet. The detachable line cord should be connected to
the IEC320 connector on the power entry module.
4.5.4

Fuses

The correct fuses must be installed into the fuse holder that is part of the AC
power entry module. Please check the fuse label on the rear panel, before
installing new fuses (see Figure 33).
Figure 33
To avoid electric shock, the appropriate fuses for the AC input
power voltage must be installed in the instrument. Only
qualified service personnel should replace fuses. Failure to
observe these precautions can result in fire, severe injury or
death.
Rear Panel
WARNING
63