Menu Section; Control Knob; Setup / Enter; Esc - Newport 6100 User Manual

Laser diode and temperature controller
Table of Contents

Advertisement

System Operation
Mode
Constant
Current
Constant
Temperature
Constant
Resistance /
Reference
4.4

Menu Section

In addition to displaying status parameters on the two 5-segment LED
displays(bottom sections) of the instrument, the Menu (top section) section of
the Model 6100 front panel enables users to view/change many more
parameters including feedback sensor type, PID values, and Steinhart-Hart or
Callendar – van Dusen coefficients, for example. Prior error messages can be
viewed as well.
4.4.1

Control Knob

The control knob on the right side of the front panel is used to set the value of
parameters such as Laser diode current setpoint (Ild), TEC temperature
setpoint (Ts) etc. Users can select the parameter whose value needs to be
modified using the Setup/Enter and Cursor Arrow keys, and then modify the
values with the knob. Alternately, the parameters' values can be modified
with the Arrow keys.
The knob has an acceleration algorithm that causes the rate of change of value
to increase as the knob is turned faster. Turning slowly allows for a fine
adjustment at the smallest displayed decimal place.
4.4.2

Setup / Enter

The Setup/Enter key is used to (a) invoke the Setup screen, (b) accept
parameter change, or (c) enter a lower menu level, depending on the screen
displayed.
4.4.3

Esc

The Esc key is used to (a) cancel a parameter change or (b) back up one menu
level.
Displayed Value
"k" LED
Actual output
OFF
current
Actual
OFF
temperature
Actual resistance
ON only if
/ reference
sensor is
thermistor.
43
o
"
C" LED
"A" LED
OFF
ON
ON
OFF
OFF
OFF

Advertisement

Table of Contents
loading