Power Supplies; Table 1 Voltage Selector Switch Settings And Fuse Ratings - Newport 6100 User Manual

Laser diode and temperature controller
Table of Contents

Advertisement

Getting Started
Temperature Controller. Verify the correct rated fuses are installed
according to the fuse marking on the rear panel.
3.6

Power Supplies

AC power is supplied through the rear panel power entry module connector
that provides in-line transient protection and RF filtering. The power entry
module also contains the instrument's fuses and the voltage selection switch.
Prior to plugging in the AC line cord, the user MUST configure the Model
6100 for operation on the available AC power. This is accomplished by
setting the voltage selector switch on the AC power entry module AND
verifying that the correct fuses are installed into the power entry module.
Failure to do this can result in degraded instrument performance, damage to
the Model 6100, and / or injury to the operator.
The voltage selector switch and the fuses are accessed by opening the power
entry module using a flat-bladed screwdriver. Note that the AC line cord
must be removed from the power entry module prior to attempting to open it.
With the power entry module open, the voltage selector switch setting can be
changed by removing the drum, rotating it to the desired setting, and
reinstalling the drum.
After correctly setting the voltage selection switch, the fuses installed in the
power entry module should be checked for the correct value and, if not
correct, changed.
Table 1, below, identifies the correct voltage selector switch setting and fuse
ratings for several common AC power
Input Power
100 VAC, 50 Hz
120 VAC, 60 Hz
220 VAC, 50 Hz
230 VAC, 50 Hz
240 VAC, 60 Hz
To avoid electric shock, the Model 6100 must be configured for
operation on your specific AC line voltage PRIOR TO
CONNECTING THE POWER CORD TO THE AC MAINS See the
instructions above for information
Voltage Switch Setting
Table 1
Voltage Selector Switch Settings and Fuse Ratings
WARNING
Fuses
100
2A, 250VAC, Slo-Blo
120
2A, 250VAC, Slo-Blo
220
1A, 250VAC, Slo-Blo
230
1A, 250VAC, Slo-Blo
230
1A, 250VAC, Slo-Blo
33