If, after a calibration check, the results exceed the tolerance in the
specification (or other suitable performance standard), carry out a
calibration adjustment.
5.3

CALIBRATION ADJUSTMENTS

If the instrument is operating correctly, only zero and full-scale
calibration will vary. Any excessive non-linearity or temperature
effects indicate a fault. The instrument should be returned to a
qualified service agent.
5.4

BEFORE STARTING

Use only original parts supplied by the manufacturer. To do an
accurate calibration, use the following:
Calibration equipment specified in Table 5-1.
Stable temperature environment: 21 ±1°C (70 ±2°F).
Before starting a calibration procedure, it is recommended leaving
the equipment in the calibration environment for a minimum of two
hours.
Function
Current
Voltage
Millivolts
Pressure
(P1)
IDOS
Page 87 of 105
Table 5-1
Calibration equipment
(ppm = parts per million)
Current (mA) calibrator.
Accuracy - Current measure/source,
See Table 5-2 and Table 5-3.
Volts calibrator.
Accuracy - Voltage measure/source,
See
Table 5-5 or Table 5-6.
mV calibrator.
Accuracy - Millivolts measure/source,
See Table 5-4 .
All Ranges total uncertainty of 0.025% reading or
better.
UPM only. Refer to the user manual for the IDOS UPM.