4. Check the calibration is correct:
5.9

PROCEDURES: PRESSURE INDICATOR

Do the procedure as follows:
1. Connect the instrument to the pressure standard.
2. Let the equipment get to a stable temperature (minimum: 60
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a two-point
calibration.
4. Check the calibration is correct:
From the datasheet use the values in the Accuracy column.
Page 92 of 105
Select the applicable Voltage (10V) or Current (24V)
(measure) function (See Section 5.7).
10V
24V
Check the error is within limits (See Table 5-6).
Table 5-6 Voltage (source) error limits
Source
V
10
24
Zero and +FS for gauge sensors.
Select the applicable pressure function.
Apply the following pressure values.
0, 20, 40, 60, 80, 100, 80, 60, 40, 20 (%FS).
Apply the following pressure values.
-200, -400, -600, -800, -600, -400, -200, 0 mbar.
Check the error is within limits
Calibrator
uncertainty
(V)
0.0001
0.0004
[EN] English – 109M4017 Revision -
Druck DPI612
Permitted
DPI612 error
(V)
0.0015
0.0036