Dirt Moisture Trap 100 Bar (P/N Io620-Idt622); Specification; Pressure Connections; Operation - GE Druck DPI612 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
6.2

Dirt Moisture Trap 100 bar (P/N IO620-IDT622)

6.2.1

Specification:

Maximum working pressure: 100bar / 1500psi pneumatic.
Important Notes:
This device is designed for use on the DPI612 PFlexPro100 bar
pneumatic pressure calibrator.
It should be removed from the calibrator during transit.
It should be inspected internally for dirt or moisture before and after
use.
Orientation: Use only in the vertical position.
6.2.2

Pressure connections:

Lower: To DPI 612 pFlexPro test port.
Upper: AMC quick-connect (for use with all DPI612 adaptor sets)
Materials of Construction: 316L and 303 Stainless steels, 6082
Aluminium, Nitrile seals, P.T.F.E.
6.2.3

Operation:

1. Check for moisture or dirt particles within the aluminium
chamber before and after use.
If found, dismantle and clean as instruction below.
2. Fit the Dirt Trap to the Test Station of the DPI 612 pFlexPro
pressure calibrator.
Hand-tight only – no spanners / wrenches are required.
3. Fit AMC adaptor to suit item under test. ( hand-tight only)
Page 96 of 105
[EN] English – 109M4017 Revision -
Druck DPI612

Advertisement

Table of Contents