Setting A Pressure Relief Valve - GE Druck DPI612 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
Hydraulic
In its normal condition, the DPI 612 hFlexPro contains hydraulic fluid. To
make sure it does not spill out, seal the system and put it on its side before
Step
1
2
3
4
2.1.3

Setting a Pressure Relief Valve

The PRV is set at the factory to operate at the maximum pressure
specified on the label (on the plastic cap). For the adjustable range,
refer to data sheet.
If necessary, use these steps to adjust the relief pressure:
Step
1
2
3
Page 41 of 105
you install a PRV.
1
Seal the system.
Put the instrument on its side.
Remove the blanking plug or, if applicable, the PRV you
are using. To collect possible drops of hydraulic fluid, put
it in a container.
Note: Before you put it into storage, make sure it is clean
and dry.
Choose a clean, dry PRV with the correct pressure value
for the devices you are using and tighten it into position
(hand tight only).
Attach an applicable PM 620 module.
Remove the plastic cap from the end of the PRV
Set the necessary pressure with the pressure station
2
Procedure
Procedure
[EN] English – 109M4017 Revision -
Druck DPI612
3
4

Advertisement

Table of Contents
loading