Procedures: Current (Source) - GE Druck DPI612 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
Apply the values that follow:
mA: -55, -25, -20, -10, -5, 0 (open circuit)
mA: 0, 5, 10, 20, 25, 55.
Check the error is in the specified limits (See Table 5-2).
Applied
0 (open circuit)
5.6

PROCEDURES: CURRENT (source)

When recalibrating measure and source functions for this range, the
measure function must be adjusted before the source function.
Do the procedure as follows:
1. Connect the applicable calibration equipment
(See Table 5-1):
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4 ) to do a two-point
calibration (0.2 mA and FS):
4. Check the calibration is correct:
Page 89 of 105
Table 5-2 Current (measure) error limits
mA
±55
±25
±20
±10
±5
CH1 (24 mA range): (See Table 5-3).
CH1: 24 mA.
Select the applicable Current (source) function
Apply the values that follow:
mA:
0.2, 6, 12, 18, 24
Check the error is within limits (See Table 5-3)
Calibrator
uncertainty
(mA)
0.0030
0.0025
0.00063
0.00036
0.00025
0.0002
[EN] English – 109M4017 Revision -
Druck DPI612
Permitted
DPI612 error (mA)
0.0057
0.0043
0.0022
0.0016
0.0013
0.0010