• A blinking LED at the bottom of the
— This display also shows the time remaining in the run (or  and elapsed time for
WARNING
Do not attempt to override the door interlock system while the rotor is spinning.
CAUTION
Do not lift or move the centrifuge while the rotor is spinning. Do not place items
on the centrifuge during operation.

Pulse Function

The pulse function, accessed by pressing the
• When the
continues to spin as long as the
deceleration settings are overridden by the pulse function.) When the
the rotor begins decelerating to 0 rpm using maximum deceleration. Using the pulse function
eliminates the need to press the
• When the
the
PULSE
to show the user-entered settings during a pulse run even though the maximum rates are used.
• If the
PULSE
at speed until the
decelerating to 0 rpm using maximum deceleration.
• The centrifuge memory retains the parameters of the last run performed before the
was pressed. At the end of a pulse run, after the centrifuge door is opened and closed the
previous run parameters are displayed.

Changing Parameters During a Run

While a run is in progress, run parameters (speed, time, temperature, and acceleration or
deceleration rate) can be altered without stopping the run. Run duration can also be changed from
continuous to a specified time period, or from a specified time period to continuous.
NOTE
The deceleration rate cannot be changed after deceleration starts.
Use the program keys as described under Entering Run Parameters, above, to change parameters.
Parameter changes made during a run must be verified by pressing the
change run speed during centrifugation:
PN B05529AE
continuous operation).
key is pressed, the rotor accelerates at maximum rate to the set speed and
PULSE
PULSE
START
key is pressed, the
PULSE
key is released, the seconds stop accumulating. The
key is pressed while the rotor is running at set speed, the rotor continues running
key is released. When the
PULSE
display indicates that the run is in progress.
TIME
key, is used for short-duration runs as follows.
PULSE
key is pressed. (The current run time, acceleration, and
and
keys.
STOP
display begins displaying the elapsed seconds. When
TIME
key is released, the rotor begins
PULSE
Operation
Run Procedure
key is released,
PULSE
display continues
ACCEL/DECEL
PULSE
key. For example, to
ENTER
2