Dc Voltage Sources 10 V And 60 V - HBM QuantumX MX840 Operating Manual

Table of Contents

Advertisement

126
6.11.13

DC voltage sources 10 V and 60 V

Supported by modules: MX840, MX840A, MX410, MX440A
Voltage ranges supported by the amplifiers:
Pin assignment for the MX1601 module see chapter 6.9.1
MX840, MX840A und MX440A:
Two measuring ranges (10 V or 60 V) can be selected, depending on the parameterization.
Wrong parameterization does not result in destruction of the amplifier.
HBM
10 V and 60 V: MX840, MX840A, MX440A
10 V: MX410, MX1601, MX1615
Maximum input voltage to housing and signal
ground :
MX840, MX840A, MX440A: "60 V
MX1601: "40 V
(+)
U
Supply voltage zero
(−)
Cable shield
1
2 Data
3 No function
Adjustable sensor supply:
1 2 3
1-wire EEPROM
(optional)
(view from below)
5 V ... 24 V:
0 V:
MX1615
Push in
6
1
1
11
5
15
10
8
14
6
4
4
5
9
7
8
Hsg.
6
1
1
Hsg.=Housing
Pin 12
Pin 11
Connection
MX1601
Push in
1
8
1
5
2
8
4
Pin 6
Pin 7
QuantumX