Safety
Electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic emissions.
Emissions test
RF emission
EN55011
RF emission
EN55011
Harmonic emission
EN 61000-3-2
Voltage fluctuations/flicker
emissions
EN/IEC 61000-3-3
Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity.
The Vscan Extend is intended for use in the electromagnetic environment below. The
customer or the user of the Vscan Extend should assure that it is used in such an
Immunity test
Electrostatic discharge
(ESD)
EN 61000-4-2
Electrical fast transient/
burst
EN 61000-4-4
Surge
EN 61000-4-5
2-14
Table 2-1:
Electromagnetic emissions
Compliance
Group 1
Class B
Group 1
Class B
Complies
Complies
Table 2-2:
Electromagnetic immunity (Part 1)
environment.
EN 60601
test level
±6 kV contact
±8 kV air
±2 kV for power-supply
lines
±1 kV for input/output
lines
±1 kV line(s) to line(s)
Electromagnetic environment - guidance
The Vscan Extend uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
The Vscan Extend is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Compliance level
±6 kV
±8 kV
±2 kV for
power-supply lines
±1 kV for input/
output lines
±1 kV line(s) to
line(s)
Vscan Extend
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
–
User Manual
5721203-100 Rev. D