Electromagnetic Compatibility

Changes or modification to this system not expressly approved by GE Healthcare
could cause EMC issues with this or other equipment. This system is designed and
tested to comply with applicable regulation regarding EMC and needs to be installed
and put into service according to the following EMC information.
WARNING:
Use of portable phones or other radio frequency (RF) emitting equipment near the
system may cause unexpected or adverse operation.
WARNING:
The equipment or system should not be used adjacent to, or stacked with, other
equipment. If adjacent or stacked use is necessary, the equipment or system
should be tested to verify normal operation in the configuration in which it is
being used.
Guidance and Manufacturer's Declaration -
Electromagnetic Emissions
The device described in this manual is intended for use in the following electromagnetic
environment. It is the responsibility of the customer or user to ensure that the device
is used in such an environment.
Emission Test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic Emissions
EN 61000-3-2
Voltage
fluctuations/Flicker
emissions
EN 61000-3-3
2046275-019H
Compliance
Electromagnetic Environment — Guidance
Group 1
The equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
Class A
Class A Equipment is suitable for use in all
Class A
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
Complies
used for domestic purposes.
MAC™ 3500
C
145