Vscan Extend Maintenance

Cleaning the device

CAUTION
Reprocessing recommendation (Frequency)
After Each Use
WARNING
6-4

Cleaning and disinfection

Make sure the device is disconnected from the AC/DC adapter
before cleaning.
1. Power off the Vscan Extend.
2. Moisten a soft, non-abrasive cloth with a mild, general
purpose, non-abrasive soap and water solution.
3. Wipe the Vscan Extend.
4. Wipe dry with a soft towel.
Do not spray the soap and water solution directly onto the
Vscan Extend.
Inspect the transducer (see 'Inspecting the Vscan Extend'
on page 6-3)
Clean the transducer (see 'Disinfecting the probe' on
page 6-9).
If required, clean the device and the display (see
'Disinfecting the device' on page 6-7).
If required, disinfect the transducer (see 'Disinfecting the
probe' on page 6-9).
Ensure that the main unit and the transducer are properly
cleaned after each use and before storage in the case.
If any defects or damages are found on the probe or its cable,
DO NOT use the Vscan Extend. Contact GE service.