DAC SETTINGS
Setting
Values
DAC
ON
AUDIBLE
OFF
DRAW
ON
OFF
MEAS
dB
%FSH
%DAC
CURVE
DAC, -2, -6,
-10, (-6/-12),
(-6/-14) dB
TRIGGER
DAC, -2, -6,
-10, (-6/-12),
(-6/-14) dB
ADJUSTING DAC SETTINGS
1. Select MENU/DAC then the desired function
2. Use LEFT and RIGHT to scroll the value until the correct value is being displayed.
3. Press OK to set the value and return to the menu screen, or ESC to cancel.
4. Press MEAS to return to the measurement screen.
Function
Enables the DAC function.
Enables audible signal.
Disables the DAC function.
When disabled, all the other menu items are greyed out.
Enables draw mode.
Disables draw mode.
Sets the quantity to be displayed when in the DAC function is
enabled:
dB - decibels
%FSH - The top of the waveform display represents 100%
full screen height. Any waveform amplitudes generated
greater than 100% full screen height will not be displayed.
100% is the greatest value that can be displayed.
%DAC - All the DAC curve represents 100%. Any signal
generating an amplitude below the DAC curve is less than
100%, and any above greater than 100%.
Note: The MEAS option selected is overridden with %FSH
when DRAW is set to ON.
Note: This option can be changed at any time after the DAC
curve has been created.
Selects curve type. DAC or DAC less a number of decibels
Note: This option can be changed at any time after the DAC
curve has been created.
Select which curve or gate to use as the alarm trigger.
Any of the DAC curves or gate displayed can be set as the
trigger alarm option.
The TRIGGER values available vary according to which
CURVE type is selected.
All the DAC curve or gate combinations only detect when the
amplitude of the signal is at or above the selected trigger
option.
Note: This option can be changed at any time after the DAC
curve has been created.
40