Elcometer FD700DL+ Operating Instructions Manual page 12

Corrosion and flaw gauge
Table of Contents

Advertisement

D
Units
E
Feature Status Bar
F
Measurement View Area RF, RECT, B-SCAN or DIGITS view area.
G
Measurement Labels
Bar
H
Hot Menu
6.2 RF A-SCAN MEASUREMENT
SCREEN
The RF view shows the full sound wave as
received by the gauge. The vertical axis
indicates the amplitude of the received wave
and the horizontal position indicates the time
it is received, which is converted into
thickness units using the material velocity.
The point on the scan that is being used to
calculate the digital thickness reading is
displayed as the vertical dashed line (see
section 9 on gates). The RF screen is used
for gauge setup prior to inspection.
6.3 RECT A-SCAN MEASUREMENT
SCREEN
The rectified view shows only half of the
waveform, either the positive or negative half
depending on the POLARITY selected.
The RECT display is the preferred view for
flaw and pit inspections
11
The current measurement units (Metric/English).
Indicates the features currently enabled and in use:
Measurement Mode (P-E, PECT, PETP, E-E, E-EV, CT)
Differential Mode (ON/OFF)
High Speed Scan Mode (ON/OFF)
Alarm Mode (ON/OFF/AUDIBLE)
Gain Setting (Displays the set value)
See "Measurement - Modes" on page 15 etc for further
information
Displays the viewable range and are based on the values of
DELAY/B-START and RANGE/B-DEPTH
The hot menu contains a selection of the most commonly used
options, allowing you to conveniently adjust these to your needs
while the graphical display is active
Hot Menu Enabled
Hot Menu Disabled