Elcometer FD700DL+ Operating Instructions Manual page 20

Corrosion and flaw gauge
Table of Contents

Advertisement

5. Press OK to select the material type and display the menu items with the new material type
selected
6. Press MEAS to return to the measurement screen.
KNOWN VELOCITY CALIBRATION
If the material velocity is known, you can enter the velocity value directly into the gauge. A list of the
sound velocities of common materials is given at the end of this instruction manual.
1. Press MENU/CAL/VELOCITY .
2. Press ENTER to display the Digits Edit Box.
3. Adjust to match the material velocity.
4. Press OK.
5. Press MEAS to return to the measurement screen.
KNOWN THICKNESS CALIBRATION
If the sound velocity of a material is unknown, a sample with one or two known thicknesses can be
used to determine the sound velocity.
Note: Although the gauge has a through-paint/coating feature, known thickness calibration must be
performed on material with the paint or coating removed.
O
-
C
NE
POINT
ALIBRATION
ranges. You should always calibrate on high side of the intended measurement range. For example,
if the measurement range is 2.54 mm to 25.4 mm (.100" to 1.0"), you should calibrate on a known
thickness sample close to 25.4 mm (1.0").
Note: Before you start this calibration procedure, perform a probe zero.
1. Apply a drop of couplant on the transducer and place the transducer in steady contact with the
sample or the material being tested. Be sure that the reading is stable and the repeatability
indicator, in the top left corner of the display, is fully lit and stable.
2. Select MENU/CAL/MATLIPT.
3. Press ENTER to display the Digits Edit Box.
4. Adjust to match the material thickness.
5. Press OK to calculate the velocity and return to the measurement screen, or ESC to cancel.
Note: Place the transducer back on the calibration point. The thickness reading should now match
the known thickness but if outside tolerance, repeat the steps above.
T
-
C
WO
POINT
ALIBRATION
one-point calibration followed by a two-point calibration. For example, if the measurement range is
2.03 mm to 6.35 mm (.080" to 0.250"), perform a one point calibration on a known thickness sample
close to 6.35 mm (.250"), followed by a two-point calibration close to 2.03 mm (.080").
Note: Before you start this calibration procedure, perform a probe zero.
Two - Point calibration
1. Conduct a one point calibration routine as described previously in this section. You should
conduct this routine at the high end of the measurement range
2. Repeat the routine with another sample at the low end of the measurement range, except this
time enter the thickness into the MATL 2PT area in the CAL menu.
Note: CHECK YOUR CALIBRATION. Place the transducer back on both calibration points. If the
thickness is out of tolerance repeat the steps above.
19
- The one point calibration option is most suited for linearity over large
- For improved accuracy over a smaller measurement range, conduct

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fd700+

Table of Contents