Transducer Sensitivity; Figure 93 - Wbv Transducer Sensitivity Selection - Castle group Excieo GA2005H Operating Manual

Human & environmental vibration meter
Table of Contents

Advertisement

Transducer Sensitivity

The Excieo vibration meter can be used with two seat pad transducers for
Whole Body Vibration, both of which are available from Castle Group Ltd.
Use this option to select the sensitivity of the seat pad being used. Below are
the sensitivities of the available seat pads:-
Whole Body Vibration Transducer Sensitivity Values
KD1009
KD1007
To ensure meter readings are correct it is essential that the sensitivity selected
matches the transducer being used with the instrument.
sensitivity setting and transducer will give incorrect measurement readings.
Press 1 for a sensitivity of 100mV/g.
Press 2 for a sensitivity of 1000mV/g.
If the sensitivity is changed then a screen prompt informs you to re-calibrate
your instrument to the new accelerometer being used. Confirmation of having
accepted that the new accelerometer must be calibrated with the new settings
of the instrument is achieved by pressing the OK button.
Selection of the change then has immediate effect in Real Time Mode and is
applied to any subsequent recordings taken in this mode.
26:08:11
ACCELEROMETER
1. 100 mV/g
2. 1000 mV/g
Select Sensitivity
CANCEL
Figure 93 – WBV Transducer Sensitivity Selection
Press Soft Key 1 labelled CANCEL to return to the previous screen without
making any changes.
09:46:15
Page 80
1000mV/g
100mV/g
26:08:11
ACCELEROMETER
1. 10 mV/g
Please re-calibrate
2. 100 mV/g
the meter for the
new accelerometer
Press OK to continue
Select Sensitivity
A mismatched
09:46:20