Castle group Excieo GA2005H Operating Manual page 125

Human & environmental vibration meter
Table of Contents

Advertisement

Measured electrically on the 1000 m/s² range of the X-axis with a reference
frequency of 80 Hz at an approximate level of 900 m/s with Fl (LIN) weighting
applied. All errors and measurements are in dB.
Frequency
Wh
(Hz)
1
-0.3
2
0.1
4
0.1
8
0.1
10
0.1
20
0.0
40
0.0
80
-0.1
100
0.0
200
0.1
400
0.0
1000
0.0
2000
-1.2
4000
-6.0
Measured electrically on the 1000 m/s² range of the X-axis with a reference
frequency of 80 Hz on Fl (LIN) for Fl, Fb filters and 8 Hz on Fl (LIN) for Fa, Fk
filters, all with an approximate level of 900 m/s. All errors and measurements
are in dB. No tolerance requirements exist for Fl or Fk.
Frequency
Fb
(Hz)
0.1
-3.4
0.2
-0.9
0.4
-0.2
1.0
0.0
2.0
0.1
4.0
0.1
8.0
0.1
10
0.1
20
0.0
40
0.1
80
0.0
100
0.0
200
0.1
400
0.1
1000
0.1
2000
-0.7
4000
-4.0
Fc
Tolerance
0.0
+2, -∞
0.1
+2, -∞
0.2
+2, -∞
0.1
±2
0.1
±1
0.0
±1
0.1
±1
0.0
±1
0.0
±1
0.1
±1
0.1
±2
0.1
±2
-0.7
+2, -∞
-4.0
+2, -∞
Tolerance
+2, -∞
-3.5
+2, -∞
-1.0
±1
-0.3
±1
-0.1
±1
0.0
±1
0.0
±1
0.0
±1
0.0
±1
0.0
±1
0.0
±1
-0.1
±1
-0.1
±1
0.0
±1
-0.1
±2
±2
+2, -∞
Page 106
Fa
Tolerance
+2, -∞
+2, -∞
±1
±1
±1
±1
±1
±1
±1
±1
±1
±1
+2, -∞
+2, -∞
Ws
Tolerance
2.0
+4, -∞
0.3
+4, -∞
0.1
+4, -∞
-1.0
+1 / -2
-0.6
±1
0.3
±1
0.1
±1
0.0
±1
0.0
±1
0.1
±1
0.3
±1
0.2
+1 / -2
1.3
±4
-1.2
±4
Fl
Fk
-3.0
-2.9
-0.9
-0.9
-0.3
-0.2
-0.1
0.0
0.0
0.1
-0.1
0.1
-1.4
0.1
-2.9
0.1
-17.6
0.0
-38.8
0.1
REF
-43.6
-43.6
0.0
-43.6
0.0
-43.6
0.1
0.1
0.2
0.2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Excieo ga2005sExcieo ga2005m

Table of Contents