Trouble Shooting Guide - Castle group Excieo GA2005H Operating Manual

Human & environmental vibration meter
Table of Contents

Advertisement

Trouble Shooting Guide

Question
The instrument will not power on.
The Instrument is not responding on
all or individual axis to inputted
vibration levels.
The
instrument
unexpected readings in one or more
modes of operation.
The overload indicator is permanently
on.
The overload indicator comes on
before 5% over the top of the range is
reached.
is
displaying
Page 125
Answer
Check condition of batteries and
replace if required.
If batteries are inserted and are of
adequate charge, remove one
battery
for
approximately
seconds and replace.
Turn the instrument Off, wait 10
seconds to allow the instrument to
reset and then turn back On.
Check the cable for possible
damage or incorrect attachment
to the transducer or instrument.
Ensure correct Axis coefficients
are
entered
in
Calibration screen for the mode of
operation.
Ensure the transducer is fitted
securely to the source of vibration.
Ensure cable is securely fastened.
Change to a higher range.
The overload indicator will also be
shown if the input amplifier is
saturated. This is possible even if
the displayed vibration reading on
your instrument is below the top of
the selected range because of the
applied frequency weighting filter.
10
the
Manual