7 Troubleshooting

7.1 Troubleshooting
SYMPTOMS
Patient lift feels loose.
Casters/Brakes noisy or stiff.
Noisy or dry sound from pivots.
Electric actuator fails to lift or legs fail to
open when button is pressed.
1195053-B
FAULTS
Mast/Base joint loose.
Lint, dirt, debris or the like in bearings.
Needs lubrication.
Pendant or actuator connector loose.
Battery low.
RED emergency stop button pressed IN.
Battery not connected properly to
control box.
The connecting terminals are damaged.
SOLUTION
Refer to the Assembling the Mast to the
Base section in the manual.
Refer to Replacing Front Casters and
Replacing Rear Casters sections in the
manual.
Refer to Lubricating the Lift section in
the manual.
Connect pendant or actuator connector.
Ensure connectors are seated properly
and fully connected.
Charge batteries. Refer to Charging the
Battery section in the manual.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control
box. Refer to Charging the Battery
section in the manual.
Replace the battery pack. Refer to
Charging the Battery section in the
manual.