Invacare Jasmine User Manual page 64

Hide thumbs Also See for Jasmine:
Table of Contents

Advertisement

Invacare® Jasmine™
Guidance and manufacturer´s declaration – electromagnetic immunity
The patient lift is intended for use in the electromagnetic environment specified below. The customer or the user of the patient
lift should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power supply
Electrostatic
lines
transient / burst
± 1 kV for input/output
IEC 61000-4-4
lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
< 5% U
for 0,5 cycle
Voltage dips, short
interruptions and
40% U
voltage variations
for 5 cycles
on power supply
70% U
input lines
for 25 cycles
IEC 61000-4-11
< 5% U
for 5 sec
64
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
(>95% dip in U
)
< 5% U
T
T
for 0,5 cycle
(60% dip in U
)
40% U
T
T
)for 5 cycles
(30% dip in U
)
70% U
T
T
for 25 cycles
(>95% dip in U
)
< 5% U
T
T
for 5 sec
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Product is double-insulated. There are no other
possible connections to earth
(>95% dip in U
)
T
T
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
(60% dip in U
the patient lift requires continued operation during
T
T
power mains interruptions, it is recommended
that the patient lift be powered from an
(30% dip in U
)
T
T
un-interruptible power supply or a battery.
U
is the a. c. mains voltage prior to application
T
(>95% dip in U
)
T
T
of the test level.
1195053-B

Advertisement