Performing An Emergency Stop - Invacare Jasmine User Manual

Hide thumbs Also See for Jasmine:
Table of Contents

Advertisement

It is recommended that the primary emergency
release be used. The secondary emergency release
is only a back-up to the primary emergency release.
The secondary emergency release is used only to
lower the patient.
In cases where the primary release is either not
functioning or unreachable, a secondary emergency
release may be used.
The secondary release will only operate when there
is a patient in the lift.
The secondary emergency release is set for 75 kg
(165 lbs). If the patient weight is more or less
than 75 kg (165 lbs), the speed of the secondary
emergency release must be adjusted.
1. Adjust the speed of the secondary release:
a. Increase the speed — Use a screwdriver to loosen
the screw in the EMERGENCY grip A.
b. Decrease the speed — Use a screwdriver to tighten
the screw in the EMERGENCY grip.
2. Pull up on the EMERGENCY grip A and push down on
the boom B at the same time to lower the patient.
1195053-B

5.6 Performing an Emergency Stop

1. Press the RED button A on the control box B in to stop
the boom and patient from raising or lowering.
2. To reset, rotate the emergency button clockwise.
Usage
23

Advertisement

Table of Contents