Invacare Jasmine User Manual page 44

Hide thumbs Also See for Jasmine:
Table of Contents

Advertisement

Invacare® Jasmine™
SYMPTOMS
Unusual noise from actuator.
Boom will not lower in uppermost
position.
Boom will not lower during a power
retraction.
If problems are not remedied by the suggested means, please contact your dealer or Invacare.
44
FAULTS
Boom or leg actuator in need of service
or load is too high.
Actuator is worn or damaged or spindle
is bent.
Boom requires a minimum weight load
to lower from the uppermost position.
Shoulder bolt at the junction of the
boom and mast may not be properly
installed.
SOLUTION
Refer to Replacing the Electric Actuator
section or Replacing the Leg Actuators
section in the manual. Contact your
Dealer or Invacare representative.
Refer to Replacing the Electric Actuator
section or Replacing the Leg Actuators
section in the manual. Contact your
Dealer or Invacare representative.
Pull down slightly on the boom.
Refer to Checking and Tightening Mast
Pivot section in the manual.