Calibration Proceedings Example; Manual Calibration, Automatic Temperature Compensation Mode; Buffer 1 - Hach POLYMETRON 8362 User Manual

Ph measurement
Hide thumbs Also See for POLYMETRON 8362:
Table of Contents

Advertisement

Transmitter 9135 - pH measurement in pure water

Calibration proceedings example

pH CALIB.

BUFFER 1

Press
ENTER
pH CALIB.
6.98
pH
0
06 . 98 pH
Manual calibration, automatic temperature
compensation mode
Proceed as follows (refer also to Fig 6.1 on page 37):
Stop circulation of the sample.
Buffer 1
Maintain the screw nut in position and unscrew the
chamber (A).
Empty out the water contained in the chamber,
and rinse with demineralised water.
Refill with a fresh buffer (B) pH 7.00 (or 6.88 at
25°C), re-screw the nut (C).
Validate by Enter.
No calibrated pH measure is displayed.
Press OK when the measure is constant.
Note :
Press ESC to quit the calibration.
The temperature indication disappears and the pH
measure appears.
The measured temperature by means of the 8362
temperature probe system allows to adjust the
value of the buffer solution.
Utilize the
Enter. Press directly Enter to not change a digit.
After the last number, press Enter to validate the
calibration.
to change a digit and press
Page 37