Specifications - Rigol DS1074B Service Manual

Ds1000b series digital oscilloscope
Hide thumbs Also See for DS1074B:
Table of Contents

Advertisement

RIGOL

Specifications

Acquisition
Sample Mode
Sample Rate
Average
Input
Input Coupling
Input Impedance
Probe Attenuation
Factors
Maximum Input Voltage
Time delay between
channel (Typical)
Horizontal
Sample Rate Range
Waveform Interpolation
Memory Depth
Scan Speed Range
(S/div)
Sample Rate and
Delay Time Accuracy
Vertical
A/D Converter
Sensitivity (V/div)
Range
Offset Range
Equivalent Bandwidth
1-2
Real-Time
2GSa/s (half channel
1GSa/s (each channel)
After all the channels finish N samples at the same time,
N can be 2, 4, 8, 16, 32, 64, 128 and 256
DC, AC, GND
1MΩ±2.0%
The input capacity is 18pF±3pF
0.001X, 0.01X, 0.1X, 1X, 2X, 5X, 10X, 20X, 50X, 100X,
200X, 500X, 1000X
Maximum input voltage of the analog channel:
CAT I 300Vrms, 1000Vpk; transient overvoltage
1000Vpk
CAT II 100Vrms, 1000Vpk
RP2200 10:1, CAT II 300Vrms
RP3300A 10:1, CAT II 300Vrms
500ps
3.65Sa/s to 2GSa/s (Real-Time),
3.65Sa/s to 50GSa/s (Equivalent-time)
Sin(x)/x
[1]
Half channel
: 16k sample points when the horizontal
timebase is 20ns/div or lower
8k sample points when the horizontal
timebase is 50ns/div or higher
Each channel: 8k sample points
1ns/div~50s/div, DS1204B
2ns/div~50s/div, DS1104B
5ns/div~50s/div, DS1074B
1-2-5 Sequence
±50ppm (any time interval that is ≥1ms)
8bit resolution, two channels sample simultaneously
2mV/div to 10V/div at input BNC
±40V (250mV/div to 10V/div)
±2V (2mV/div to 245mV/div)
70MHz (DS1074B)
Chapter 1 Specifications
Equivalent-time
[1]
)
50GSa/s
DS1000B Service Guide
[2]

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1104bDs1204b

Table of Contents