Chapter 4 Calibration

Figure 4-2 Self-Calibration Finishes

Generally, the self-calibration will last for about 10 to 15 minutes. If the instrument
does not pass the self-calibration after this period of time or the progress bar stops at
one of the calibration items, failure might occur to the instrument. Please restart the
instrument and perform the self-calibration again; if the problem still remains, please
contact RIGOL technical support.
4-2
DS1000B Service Guide