Check the 4 ... 20 mA
signal
Error
4 ... 20 mA signal not sta-
ble
4 ... 20 mA signal missing
Current signal greater than
22 mA or less than 3.6 mA
Treatment of measure-
ment errors with liquids
VEGAPULS 62 • 4 ... 20 mA/HART - four-wire
Further comprehensive diagnostics options are available with a PC
with PACTware and the suitable DTM. In many cases, the reasons can
be determined in this way and faults rectified.
Connect a multimeter in the suitable measuring range according to
the wiring plan. The following table describes possible errors in the
current signal and helps to eliminate them:
Reason
•
Fluctuations of the
measured variable
•
Electrical connection
faulty
•
Voltage supply missing
•
Operating voltage too
low or load resistance
too high
•
Electronics module in the
sensor defective
The below tables show typical examples of application-related meas-
urement errors with liquids. The measurement errors are differentiated
according to the following:
•
Constant level
•
Filling
•
Emptying
The images in column "Error pattern" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
1
Real level
2
Level displayed by the sensor
Notes:
•
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
•
If the level indication is too low, the reason could be a line resist-
ance that is too high
9 Diagnosis, asset management and service
Removal
•
Set damping appropriate to the instrument via the
display and adjustment module or PACTware/DTM
•
Check connection according to chapter "Connection
steps" and if necessary, correct according to chapter
"Wiring plan"
•
Check cables for breaks; repair if necessary
•
Check, adapt if necessary
•
Exchange the instrument or send it in for repair
0
1
2
time
63