Vega VEGAPULS 62 Operating Instructions Manual page 63

Radar sensor for continuous level measurement of liquids
Hide thumbs Also See for VEGAPULS 62:
Table of Contents

Advertisement

Check the 4 ... 20 mA
signal
Error
4 ... 20 mA signal not sta-
ble
4 ... 20 mA signal missing
Current signal greater than
22 mA or less than 3.6 mA
Treatment of measure-
ment errors with liquids
VEGAPULS 62 • 4 ... 20 mA/HART - four-wire
Further comprehensive diagnostics options are available with a PC
with PACTware and the suitable DTM. In many cases, the reasons can
be determined in this way and faults rectified.
Connect a multimeter in the suitable measuring range according to
the wiring plan. The following table describes possible errors in the
current signal and helps to eliminate them:
Reason
Fluctuations of the
measured variable
Electrical connection
faulty
Voltage supply missing
Operating voltage too
low or load resistance
too high
Electronics module in the
sensor defective
The below tables show typical examples of application-related meas-
urement errors with liquids. The measurement errors are differentiated
according to the following:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
1
Real level
2
Level displayed by the sensor
Notes:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
If the level indication is too low, the reason could be a line resist-
ance that is too high
9 Diagnosis, asset management and service
Removal
Set damping appropriate to the instrument via the
display and adjustment module or PACTware/DTM
Check connection according to chapter "Connection
steps" and if necessary, correct according to chapter
"Wiring plan"
Check cables for breaks; repair if necessary
Check, adapt if necessary
Exchange the instrument or send it in for repair
0
1
2
time
63

Advertisement

Table of Contents
loading

Table of Contents