Vega VEGAPULS 62 Operating Instructions Manual page 22

Radar sensor for continuous level measurement of liquids
Hide thumbs Also See for VEGAPULS 62:
Table of Contents

Advertisement

4 Mounting
Surge pipe extension
Instructions and require-
ments, surge pipe
22
ø 60,3 mm
(2.37")
15 mm
(0.59")
2 mm
(0.08")
1
4 mm
(0.16")
Fig. 18: Welding connection with surge pipe extension for different example
diameters
1
Position of the welded joint with longitudinally welded pipes
Instructions of orientation of the polarisation:
Note marking of the polarisation on the sensor
With threaded versions, the marking is on the hexagon, with flange
versions between two flange holes
The marking must be in one plane with the holes in the surge pipe
Instructions for the measurement:
The 100 % point must be below the upper vent hole and the
antenna edge
The 0 % point is the end of the surge pipe
During parameter adjustment, select "Application standpipe" and
enter the tube diameter to compensate for errors due to running
time shift
A false signal suppression with the installed sensor is recom-
mended but not mandatory
The measurement through a ball valve with unrestricted channel
is possible
Constructive requirements:
Material metal, smooth inner surface
Preferably pultruded or straight beaded stainless steel tube
Welded joint should be straight and lie in one axis with the holes
Flanges are welded to the tube according to the orientation of the
polarisation
ø 88,9 mm
(3.5")
26 mm
(1.02")
2 mm
(0.08")
1
5 mm
(0.20")
VEGAPULS 62 • 4 ... 20 mA/HART - four-wire
ø 114,3 mm
(4.5")
30,5 mm
(1.20")
2 mm
(0.08")
1
5 mm
(0.20")