Installing

3.3 Installing the wall-mounted device

3.2.2
Installing
Procedure
● Fasten the rack-mounted device with suitable screws and suitable tools.
● The rack-mounted device can optionally be positioned on brackets or base plates in a
The dimensions for the drill holes are specified in the dimension drawing.
Figure 3-1
3.3
Installing the wall-mounted device
3.3.1

Notes on installation

Note
Protect the wall-mounted device from:
• Direct thermal radiation
• Mechanical damage
The installation location should be:
• easily accessible
• free from vibration
• within the ambient temperature limits
Note
Air circulation
Ensure that heat does not build up at the place of installation. Ensure that there is a sufficient
circulation of air around the wall-mounted device.
If more than one wall-mounted device is installed, there should be a lateral distance of ≥
5 cm between them.
14
frame or cabinet.
10,3
13,5
Dimension drawing
170,8
482,3