Analyzer module
These functions are permanently assigned (no parameter assignment is possible):
Table 7- 10
Analyzer module: permanently assigned digital outputs
Relay
Functions
1
1
Safe measurement
Unsafe measurement
Fault
2
Function check "On"
Function check "Off"
3
Fault
No alarm
Maintenance demanded
Maintenance required
Cannot be set
1)
7.8.3

Setting digital outputs

Procedure
1. Main menu > "2 Settings" > "09 Inputs/outputs" > "5 Digital outputs"
2. Open selected digital output: press <ENTER>.
3. Define function group: "Function (pre-selection)" parameter field.
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
Active level
Enabled
Passive
Passive
Passive
Enabled
Passive
Enabled
Enabled
Enabled
The menu includes navigation lines for access to all available digital outputs. The number
of available digital outputs depends on the hardware of your device.
Structure of the navigation line for each digital output
Navigation
DO name
no.
1) 2) 3)
Standard: Digital outputs 01 to 08.
1)
Only with option module 1: Digital outputs 09 to 20.
2)
Cannot be configured: Digital outputs 21 to 23 (analyzer module 1), digital outputs 24 to 26
3)
(analyzer module 2).
Display only for function (pre-selection) "Limit", "Solenoid valve", "Measuring range"
4)
The parameter display of a digital output whose parameters have not yet been assigned
contains the "Function (pre-selection)" and "Function" parameters fields, each with the
setting "Free".
The selection of the function group determines additional setting or display options.
Remark/cause
Digital outputs 21 or 24.
The relay serves as status relay that signals an exit from safe
measurement.
Digital outputs 22 or 25.
The relay signals whether the module is in "Function check" mode.
Digital outputs 23 or 26.
The relay is used in analyzer modules to signal the fault status.
Function
Commissioning
7.8 [2.09.5] Digital outputs
Component
4)
Edit